Story about medical power Supply

There are two main users of medical power supply: one is patient, another operator. The operator is more professional to use.  So, at IEC60601-1 3rd edition, MOP (Means of Protection) distinguished MOPP(Means of Patient Protection) and MOOP (Means of Operator Protection) with the following details:

classifications

isolation

Creepage/clearance

Insulation

1*MOOP

1500VAC

2.5mm/2mm

basic

2*MOOP

3000VAC

5mm/4mm

double

1*MOPP

1500VAC

4mm/2.5mm

basic

2*MOPP

4000VAC

8mm/5mm

double

The medical power supply manufacturer shall make sure if the patient touch this medpower. If yes, this medpower has to meet MOPP insulation.

 Whichever MOPP or MOOP, primary to secondary is 2*MOP, primary to FG (frame ground) at least 1*MOP.