Astek catch your demand on Power Supply Astek catch Your Demand on Power Supplies ' https://www.astekglobe.com/Deutsche ”https://www.astekglobe.com/sp” ”https://www.astekglobe.com/fr” ”https://www.astekglobe.com/jp” "https://www.astekglobe.com" ”https://www.astekglobe.com/cn” ” https://www.astekglobe.com/zh” ”https://www.astekglobe.com”
Home Products Medical Adapter AAM250 240W-250W Medical Desktop Adapter
222

AAM250 240W-250W Medical Desktop Adapter

*Input 90-264VAC *Output 12V/19V/24V/48V *UL/cUL60601-1,ANSI/AAMI/IEC/EN/ES 60601-1:2012 (60601-1 edition 3.1) *EMC:IEC60601-1-2:2014 (edition 4.0) *2 x MOPP *DoE VI, CoC Tier 2, NRCan,GEMS *Touch Current <100uA *Modified and Customized design available

74801d46c7fe26ab3673f786b92f53de.jpg

AAM250 240W-250W Medical Desktop Adapter

*100-240 Universal AC Input Range

*Regulated Output with low ripple noise

*Protections: Short Circuit/ Overvoltage/Over current/Over Temperature

*Touch Current <100uA

*DOE VI, Energy star, ErP step 2

*UL/cUL60601-1,ANSI/AAMI/IEC/EN ES60601-1:2012(60601 edition 3.1)

* EMC: IEC60601-1-2:2014(edition 4.0)

*Means of Protection: 2 * MOPP

*Modified and custom design provided

*2 years warranty

Input Voltage: 90~264VAC

Frequency: 50~60Hz

Input current: 3.3-1.3A

Protection: Internal Primary current fuse

Efficiency : DOE VI, Energy star, ErP step 2, CEC

Touch Current <100uA

Part No Rated O/P(current max.)

AAM250-S12-z 12V/20A/240W

AAM250-S19-z 19V/13.2A/250W

AAM250-S24-z 24V/10.42A/250W

AAM250-S48-z 48V/5.2A/250W z=C14, Size: 204*94*43.5(mm), weight: 1040g

Load Regulation: +/-5% (Typical)

Ripple 2% Vp-p Max. for output voltage @full load

Transient Response: 0.5mS for 50% Load Change Typical

Hold Up Time: 10mS @full load

Operating Temperature: 0~40degree C

Storage Temperature: -20~80degree C

Operating Humidity: 20~80% RH

Storage Humidity: 10~90% RH Topology: LLC

Dielectric withstand: 4000VAC Primary to Secondary

EMI Conduction& Radiation: Compliance EN55011 class B
Safety: UL/cUL60601-1,ANSI/AAMI/IEC/EN/ES 60601-1:2012 (60601-1 edition 3.1)
EMC:IEC60601-1-2:2014 (edition 4.0) 

Harmonic current: Compliance EN61000-3-2 EMS Immunity: Compliance IEC60601-1-2

MTBF: 300,000 Calculated Hours at 25。C, by Telcordia SR-332

Cooling: Natural Cooling convection