AAM036 25W-36W Medical Desktop Adapter
*医用电源适配器*输入电压100-240VAC +/- 10%*输出5V / 7.5V / 9V / 12V / 15V / 18V / 24V * ANSI / AAMI / IEC / EN ES60601-1:2012(60601版3.1)* EMC:IEC60601-1-2:2014(版本4.0)*保护手段:2 * MOPP
AAM036 25W-36W Medical Desktop Adapter
*100-240 Universal AC Input Range
*Regulated Output with low ripple noise
*Protections: Short Circuit/ Overvoltage/Over current/Over Temperature
*DOE VI, Energy star, ErP step 2
*ANSI/AAMI/IEC/EN ES60601-1:2012(60601 edition 3.1)
* EMC: IEC60601-1-2:2014(edition 4.0)
*Means of Protection: 2 * MOPP
*Modified and custom design provided
*2 years warranty
Input Voltage: 90~264VAC
Frequency: 50~60Hz
Input current: 1.0-0.45A
Protection: Internal Primary current fuse
Efficiency : DOE VI, Energy star, ErP step 2, CEC
Touch Current <100uA
Part No Rated O/P(current max.)
AAM036-S05-z 5V/5A/25W
AAM036-S075-z 7.5V/4A/30W
AAM036-S09-z 9V/4A/36W
AAM036-S12-z 12V/3A/36W
AAM036-S15-z 15V/2.4A/36W
AAM036-S18-z 18V/2A/36W
AAM036-S24-z 24V/1.5A/36W z=C14/06/08/18 inlet, Size: 100*50*33(mm), weight: 210g
Load Regulation: +/-5% (Typical)
Ripple 2% Vp-p Max. for output voltage @full load
Transient Response: 0.5mS for 50% Load Change Typical
Hold Up Time: 10mS @full load Topology: LLC
Operating Temperature: 0~40degree C Storage Temperature: -20~80degree C
Operating Humidity: 20~80% RH Storage Humidity: 10~90% RH
Dielectric withstand: 4000VAC Primary to Secondary
EMI Conduction& Radiation: Compliance EN55011 class B
Harmonic current: Compliance EN61000-3-2 EMS Immunity: Compliance IEC60601-1-2
MTBF: 300,000 Calculated Hours at 25。C, by Telcordia SR-332
Cooling: Natural Cooling convection